Presentation2 eflexis and erytra

Published on
Embed video
Share video
Ask about this video

Scene 1 (0s)

Chuene.

Scene 3 (10s)

Erytra/Eflexis. Erytra /Eflexis is a fully automatic analyzer designed to automate invitro immunohaematological tests on human blood, using Grifols gel card technology, including blood group and antigen typing, antibody screening, compatibility tests and direct Coombs test. Results are automatically read, interpreted and registered. Results can be reviewed, printed or exported upon user request The machine test the below: ABO Red Cell and Serum grouping. Rh(D) typing. Antigen typing. Antibody detection. Antibody identification..

Scene 4 (33s)

Erytra.

Scene 6 (6m 44s)

Troubleshooting Eflexis and Erytra. Initialization If the instrument has been shut down incorrectly , manually shut down the instrument down for 5 second and unplug from the power source and put it back on again When the instrument is on but still can’t initialise probable it is the fluids that are the issue. It normally happens if the instrument has not been active for a while. When it starts initialising you will see with the below icon. You can solve this by performing several primes and washings from the Technical Service Door opened: it is necessary to have correctly closed that the upper door and card drawers to allow the complete initialization of the instrument..

Scene 7 (7m 13s)

Initialization. [image] GRIFOLS.

Scene 8 (7m 19s)

The barcode reader from the gripper is not able to identify any barcode.

Scene 9 (7m 47s)

The barcode reader from the gripper is not able to identify any barcode.

Scene 10 (7m 53s)

Incubator Troubleshooting. The machine starts incubating the cards at around 25 to 37°C. This sometimes get disabled but you can enable them under maintenance tab. If the issue persist a technician must have a look at the Peltiers. If the module is faulty it will normally appear as red.

Scene 11 (8m 9s)

Gripper Error. [image] 0204034000 Error during the gripper robotic arm movement. Plase, inspect the instrument cards area. Please choose one of the following options: RETRY CANCEL works.

Scene 12 (8m 20s)

Gripper Error. You can retry the action again, if not solved you can cancel the work that it is trying to do. It is sometimes disabled you can enable it again Controller>Maintenance>Incubators&Centrifuges tab when the instrument is not performing any test If the issue persist call FSE, they needs to perform a teach on the gripper..

Scene 13 (8m 38s)

Sample /Reagent Detection. The presence detector of the PIP Arm cannot identify sample or reagent on the rack. If not, check that the orange light of the laser is on while it is detecting the presence of the vial or tube. If problems persist for reagents: If the system detects “ghost” vials, polish the floor of the reagents’ rack in each position where the vials are placed. If the system does not detect the presence of the real vials, check the adjustment of the laser beam. Samples: Observe the samples’ rack and check that samples are correctly placed and completely parallel between them, labels can interfere in the automatic detection of the diameter tube. Detector of the PIP Arm cannot identify sample or reagent on the rack. If not, check that the orange light of the laser is on while it is detecting the presence of the vial or tube. If problems persist for reagents: If the system detects “ghost” vials, polish the floor of the reagents’ rack in each position where the vials are placed. If the system does not detect the presence of the real vials, check the adjustment of the laser beam..